• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. 130 RAD DRILL GDE DROP; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. 130 RAD DRILL GDE DROP; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 71674003
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Date 10/08/2020
Event Type  malfunction  
Event Description
It was reported that, during inspection, button was noticed to be broken and spring is missing.No case involved.
 
Manufacturer Narrative
Internal complaint reference (b)(4).Results of investigation: the device, intended for use in treatment was returned for evaluation.Visual inspection of the returned device confirms that the device is broken and has signs of wear and tear from use.The device was manufactured in 2017.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.A review of complaint history for the listed part revealed no prior complaints.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.The device is a reusable device that can be exposed to numerous surgeries and cleaning cycles.Probable causes that could contribute to the reported event, as reusable instruments are susceptible to damage from prolonged use and through misuse or rough handling.To avoid compromised performance or damage, it is recommended the device be inspected prior to and after each use and cleaning.We recommend that all re-usable instruments be routinely inspected for wear/damage and replaced as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
130 RAD DRILL GDE DROP
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10774541
MDR Text Key214296307
Report Number1020279-2020-06026
Device Sequence Number1
Product Code HSB
UDI-Device Identifier03596010560476
UDI-Public03596010560476
Combination Product (y/n)N
PMA/PMN Number
K122170
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 11/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/02/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71674003
Device Catalogue Number71674003
Device Lot Number17CM12533
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/03/2020
Date Manufacturer Received11/16/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-