• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-85
Device Problem Pumping Problem (3016)
Patient Problem No Patient Involvement (2645)
Event Date 10/05/2020
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate the device and traced the fault back to patient pump 1.The part was replaced and the issue solved.Subsequent functional verification testing was completed without further issues and the unit was returned to service.The issue was most likely caused by environmental conditions in which the pumps operate.Water infiltration or water formation due to condensation or high temperatures and high level of humidity in the stationary phase may led to bearing damage not allowing the pump the rotate freely.
 
Event Description
Livanova (b)(4) received a report that a heater-cooler system 3t displayed two error codes associated to patient pumps during priming.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80309
GM  80309
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key10776551
MDR Text Key214314746
Report Number9611109-2020-00616
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K191402
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date 11/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/03/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-85
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/05/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/07/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-