• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR Back to Search Results
Model Number 37601
Device Problems Energy Output Problem (1431); Therapy Delivered to Incorrect Body Area (1508); Patient Device Interaction Problem (4001)
Patient Problems Undesired Nerve Stimulation (1980); Device Overstimulation of Tissue (1991); Twitching (2172); Complaint, Ill-Defined (2331); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Type  malfunction  
Manufacturer Narrative
The event date is unknown.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported the patient's legs are getting worse and their face has started to twitch.They stated when they look down at the tv, the twitching goes away, but when they look up, it pulls on their cheeks a bit (stimulation), and makes them twitch and go up.They connected to their implantable neurostimulator (ins) and everything was on and ok (2.90a, 3.20 are their current settings).They will contact their healthcare provider (hcp) to get direction on if they should change the settings or go in for an appointment.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information received from the consumer reported the cause of the stimulation in the cheeks, face twitching, and legs worsening was due to the battery voltage being too high.The battery ¿voltage¿ was lowered and the problem was solved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
MDR Report Key10784893
MDR Text Key214547968
Report Number3004209178-2020-19280
Device Sequence Number1
Product Code MHY
UDI-Device Identifier00763000175719
UDI-Public00763000175719
Combination Product (y/n)N
PMA/PMN Number
P960009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 01/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/04/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/28/2020
Device Model Number37601
Device Catalogue Number37601
Was Device Available for Evaluation? No
Date Manufacturer Received01/20/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-