Model Number HST III SEAL (4.5MM), 5-PACK
Device Problem Activation Problem (4042)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/11/2020
Event Type  malfunction  
Manufacturer Narrative
Trackwise id #: (b)(4). Since the device is not available to be returned to us, a technical evaluation cannot be performed. Per our standard sop's, all events are tracked and trended to determine whether or not any trends develop. Device discarded.
 
Event Description
The hospital reported that during a coronary artery bypass procedure using hst iii seal (4. 5mm), there was a problem with the deployment of the proxima seal when using heart string iii. The delivery operation of the delivery system was performed on the delivery system (it was normal), and after inserting the delivery system into the aorta, the delivery system was removed, but it could not be removed normally. Therefore, the aorta opening could not be removed normally. The part could not be completely closed by the delivery system. They decided that there was a problem and removed the proxima centauri, while holding down with their finger to stop bleeding, another heart string iii (hs-3045) was used. The second use was able to perform central anastomosis without any problem. The hospital did not report any patient effects.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameHST III SEAL (4.5MM), 5-PACK
Type of DeviceCLAMP, VASCULAR
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ 07470
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ 07470
Manufacturer Contact
dorota wolpiuk
45 barbour pond drive
wayne, NJ 07470
MDR Report Key10785793
MDR Text Key214582372
Report Number2242352-2020-00944
Device Sequence Number1
Product Code DXC
UDI-Device Identifier00607567700307
UDI-Public00607567700307
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K130382
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial,Followup
Report Date 11/02/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received11/04/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date02/10/2021
Device Model NumberHST III SEAL (4.5MM), 5-PACK
Device Catalogue NumberHS-3045
Device Lot Number25150169
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Event Location No Information
Date Manufacturer Received04/26/2021
Was Device Evaluated by Manufacturer? No
Date Device Manufactured02/11/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-