Model Number CI-1400-01 |
Device Problems
Expulsion (2933); Material Protrusion/Extrusion (2979); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 09/16/2020 |
Event Type
Injury
|
Event Description
|
The recipient reportedly experienced device extrusion.The recipient's device was explanted.Advanced bionics is in the process of obtaining additional information.When additional information is received, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
The recipient was reportedly not given any additional treatment prior to explantation surgery.The recipient was not re-implanted.The recipient is doing well.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed the electrode was severed prior to receipt.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The device passed the electrical and mechanical tests performed.This device was explanted for medical reasons.The device passed the tests performed.This version of the hires device is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|