• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG, PRECISION HD 30 DEG., ARTHROSCOPE, 4.0MM X 140MM, C-MOUNT, SPEEDLOCK, A/C

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG, PRECISION HD 30 DEG., ARTHROSCOPE, 4.0MM X 140MM, C-MOUNT, SPEEDLOCK, A/C Back to Search Results
Model Number 0502904030
Device Problems Poor Quality Image (1408); Optical Problem (3001)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/09/2020
Event Type  malfunction  
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event Description
It was reported that there was a blurry image.
 
Manufacturer Narrative
Alleged failure: image not clear.Confirmed failure: outer tube damaged (bent, dented),loose prism,damaged distal cover glass.Probable root cause: incorrectly assembled optical train.Damage to optical train.Debris in optical train.End of life wear-out.Shipping damage.Use error.The device manufacturer date is not known.The product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.
 
Event Description
It was reported that there was a blurry image.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG, PRECISION HD 30 DEG., ARTHROSCOPE, 4.0MM X 140MM, C-MOUNT, SPEEDLOCK, A/C
Type of Device
ARTHROSCOPE
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key10795180
MDR Text Key217704859
Report Number0002936485-2020-00459
Device Sequence Number1
Product Code HRX
UDI-Device Identifier07613327063844
UDI-Public07613327063844
Combination Product (y/n)N
PMA/PMN Number
K093677
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 12/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/05/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0502904030
Device Catalogue Number0502904030
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/21/2020
Date Manufacturer Received10/09/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-