Model Number 71674523
Device Problem Mechanical Problem (1384)
Patient Problems No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/07/2020
Event Type  Injury  
Event Description
It was reported that the lag screw drill sleeve was extremely difficult to remove from the 3.2 guide pin sleeve during surgery.The surgeon had to use other tools to remove it prior to drilling with the lag screw drill.No surgical delay or injury to the patient was reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
The device, used treatment, was returned for evaluation.A visual inspection of the returned device confirms significant signs of wear and usage.There are some scratches, burrs and gouges in the device.This device was manufactured in 2014.A functional evaluation was conducted with the mating device returned with this device.The pin sleeve was not able to target correctly.A medical investigation was conducted and this case reports that during the procedure, the lag screw drill sleeve was extremely difficult to remove from the guide pin sleeve.Per email communication, different instrumentation was used to remove the guide pin sleeve from the lag screw sleeve, prior to drilling with the lag screw drill.There was no patient injury or delay reported.No further clinical assessment is warranted at this time.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of risk management files found that the reported failure was documented appropriately.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAG SCREW DRILL SLEEVE
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10795793
MDR Text Key214989633
Report Number1020279-2020-06244
Device Sequence Number1
Product Code HSB
UDI-Device Identifier00885556363249
UDI-Public00885556363249
Combination Product (y/n)N
PMA/PMN Number
K122170
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 12/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/05/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71674523
Device Catalogue Number71674523
Device Lot Number14JCT0001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/10/2020
Date Manufacturer Received12/09/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-