Device Problem Gas Leak (2946)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/10/2020
Event Type  malfunction  
Manufacturer Narrative
A steris service technician arrived onsite to inspect the unit and found one of the contactors for their steam generator was stuck in the closed position. As the contactor was stuck in the closed position, this allowed continuous power to be provided to the unit causing excess steam to build up. As designed, the safety valve opened, and excess steam emitted from the unit. The user facility has been provided a quote for a replacement generator. No additional issues have been reported.
 
Event Description
The user facility reported steam was leaking from their amsco 400 sterilizer. No report of injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameAMSCO 400 STERILIZER
Type of DeviceSTERILIZER
Manufacturer (Section D)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX 67190
Manufacturer (Section G)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX 67190
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
4403927453
MDR Report Key10800384
MDR Text Key219524447
Report Number3005899764-2020-00076
Device Sequence Number1
Product Code FLE
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/06/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received11/06/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received10/10/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-