(b)(6).Device evaluated by mfr.: the device was returned for analysis.The shaft, hypotube, tip and balloon were microscopically and visually examined.Upon visual and microscope examination of the shaft, hypotube, tip and balloon.There were numerous kinks noted.There was a complete separation at 81.2cm distal of the strain relief.There was contrast in the inflation lumen and balloon.The balloon was loosely folded.Microscopic inspection revealed tip damage.Inspection of the remainder of the device presented no other damage or irregularities.
|
Reportable based on device analysis completed on 16oct2020.It was reported that a shaft kinked had occurred.The 90% stenosed lesion was severely tortuosed and severely calcified.An 3.00mm x 8.00mm nc emerge balloon was selected for use.During the procedure, it was noted that the shaft was kinked.The procedure was completed with a different device.No patient complications were reported.However, device analysis noted a complete separation at 81.2cm distal of the strain relief.
|