Catalog Number 8065751658
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/03/2020
Event Type  malfunction  
Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A physician reported a leakage from the trocar before surgery, the product was replaced and the procedure was completed.There was no patient harm.
 
Manufacturer Narrative
A sample was not received at the manufacturing site for evaluation for the report of trocar leakage; therefore, the condition of the product could not be verified.A review of the device history record traceable to the reported lot number, indicates that the product was processed and released according to the product¿s acceptance criteria.Because a sample was not received at the manufacturing site and the device history record review indicated that the product was processed and released according to the product¿s acceptance criteria, the exact root cause for this complaint is unknown.Investigations have been completed and manufacturing process enhancements have been implemented in order to improve the performance of the valves.Complaints are reviewed and monitored at regular intervals to evaluate effectiveness of actions taken.No additional action is required at this time.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
Three opened 25 ga trocar assemblies in a smpt were received.For the report of trocar leaking.The returned samples were visually inspected.And were found to be non-conforming with damage to the septum's.Leak testing could not be performed, due to the damage of the samples.The product was processed and released according to the product¿s acceptance criteria.The exact root cause for this complaint is unknown.The damaged condition of the valves could have contributed to the reported event.How and when the valves became damaged, cannot be determined from the evaluation performed.The exact root cause for this complaint is unknown.Therefore, specific action for this complaint cannot be taken.An acceptance quality inspection is performed to ensure product meets release acceptance criteria.This inspection includes evaluation for damaged valves.Complaints are reviewed and monitored at regular intervals to evaluate effectiveness of actions taken.No additional action is required at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDGEPLUS VALVED ENTRY SYSTEM
Type of Device
CANNULA, TROCAR, OPHTHALMIC
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
MDR Report Key10802714
MDR Text Key216850945
Report Number2028159-2020-00930
Device Sequence Number1
Product Code NGY
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Remedial Action Other
Type of Report Initial,Followup,Followup
Report Date 05/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/06/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2022
Device Catalogue Number8065751658
Device Lot Number13LFRT
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/07/2021
Date Manufacturer Received05/06/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-