DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number N/A |
Device Problems
Circuit Failure (1089); Gas/Air Leak (2946)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/14/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge field service engineer (fse) evaluated the iabp unit and was unable to reproduce the reported "gas loss error." however, in evaluating the unit, the fse discovered that the unit failed pim leak test after cycling the unit at 100 bpm with a test set up.As such, the fse replaced the safety and tidal disks, and re-tested the unit five times with passing results.The fse then performed all functional and safety checks to meet factory specifications.The iabp was then released to the customer for return to clinical service.(b)(6).
|
|
Event Description
|
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) unit had a gas loss in the iab circuit, and was swapped out.No patient harm, serious injury or adverse event was reported.
|
|
Event Description
|
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) unit had a gas loss in the iab circuit, and was swapped out.No patient harm, serious injury or adverse event was reported.
|
|