• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number N/A
Device Problems Circuit Failure (1089); Gas/Air Leak (2946)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/14/2020
Event Type  malfunction  
Manufacturer Narrative
A getinge field service engineer (fse) evaluated the iabp unit and was unable to reproduce the reported "gas loss error." however, in evaluating the unit, the fse discovered that the unit failed pim leak test after cycling the unit at 100 bpm with a test set up.As such, the fse replaced the safety and tidal disks, and re-tested the unit five times with passing results.The fse then performed all functional and safety checks to meet factory specifications.The iabp was then released to the customer for return to clinical service.(b)(6).
 
Event Description
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) unit had a gas loss in the iab circuit, and was swapped out.No patient harm, serious injury or adverse event was reported.
 
Event Description
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) unit had a gas loss in the iab circuit, and was swapped out.No patient harm, serious injury or adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
MDR Report Key10805472
MDR Text Key218956642
Report Number2249723-2020-01852
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 11/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/06/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Date Manufacturer Received11/05/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-