ADVANCED BIONICS, LLC HIRES 90K ADVANTAGE IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1500-01
Device Problems Expulsion (2933); Material Protrusion/Extrusion (2979); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Insufficient Information (4580); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 10/10/2020
Event Type  Injury  
Event Description
The recipient reportedly experienced device extrusion due to a trauma incident.The recipient's device was explanted.The recipient was not reimplanted.The recipient is in the process of healing.
 
Manufacturer Narrative
Corrections: sections b.3 & d.6b additional medical intervention was not required prior to revision surgery.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed the electrode was severed prior to receipt and missing the electrode ground ring.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The device passed the electrical and mechanical tests performed.This device was explanted for medical reasons.The device passed the tests performed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES 90K ADVANTAGE IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
MDR Report Key10806380
MDR Text Key215260604
Report Number3006556115-2020-01255
Device Sequence Number1
Product Code MCM
UDI-Device Identifier07630016815805
UDI-Public(01)07630016815805(11)161121(17)181130
Combination Product (y/n)N
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup,Followup
Report Date 10/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/08/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date11/30/2018
Device Model NumberCI-1500-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/04/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age8 YR
-
-