Model Number 283512
Device Problem Corroded (1131)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/01/2020
Event Type  malfunction  
Manufacturer Narrative
Product complaint #: (b)(4). If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate. Investigation summary: the device was received and evaluated at the service center. The reported complaint that the device does not work and beeps was confirmed. It was found that the motor did not run as it was corroded. Further, the resistance values of the keypad of the hand control set and the protective earth resistance of the motor cable were out of range and the buttons of the keypad were not functioning. The motor, motor cable and the hand control set were replaced and the device was repaired, tested and found to be fully functional. Fluid ingress into the system and contact with the motor is responsible for the corrosion of the motor. The corroded motor has a tendency to stick and not turn, hence is responsible for the issue reported by the customer. However, given the information provided, we cannot determine a definitive root cause for the defective keypad of the hand control set and the defective motor cable. The service history has been reviewed in lieu of the device history record for this device since it was previously serviced. The device was last serviced on 04/05/2011 and passed all functional testing before being returned to the customer. At this point in time, no corrective action is required, and no further action is warranted. Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field. Udi: (b)(4).
 
Event Description
It was reported via personal interaction that a fms tornado micro handpiece with buttons does not work. When connecting to the pump they heard a bip. No surgical delay or patient consequences reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMICRO TORNADO HP W HANDCONTROL
Type of DeviceLINE-POWERED ARTHROSCOPIC SHAVER SYSTEM
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6103142063
MDR Report Key10810361
MDR Text Key215345300
Report Number1221934-2020-03373
Device Sequence Number1
Product Code HRX
Combination Product (y/n)N
PMA/PMN Number
K954465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other
Type of Report Initial
Report Date 10/22/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received11/09/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number283512
Device Catalogue Number283512
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/13/2020
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received10/22/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device?
Type of Device Usage

-
-