It was reported to philips the device soft keys were not functioning properly.The device was reported to be in use, however, no direct adverse event to the patient or user was reported.A philips field service engineer (fse) evaluated the device.The reported issue was confirmed and traced to a faulty processor pca.One processor pca was returned for failure analysis.The reported allegation about the "the soft keys were not functioning properly" was not verified or duplicated during lab tests.The processor pca passed all tests during opcheck and performed as expected.Therefore, there is no fault found with this processor pca.The fse replaced the processor pca.The device passed all performance assurance tests and was placed back into use with the customer.
|