WRIGHTS LANE SYNTHES USA PRODUCTS LLC SPINDLE NUT APPARATUS, TRACTION, NON-POWERED Back to Search Results
Model Number 394.42
Device Problem Material Twisted/Bent (2981)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/14/2020
Event Type  malfunction  
Manufacturer Narrative
The device was received, the investigation could not be completed, and no conclusion could be drawn, as product is entering the complaint system. Device was used for treatment, not diagnosis. If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that on (b)(6) 2020, patient underwent an open reduction internal fixation (orif) of right tibial plateau. During the procedure the nut on the femoral distractor got stuck in the middle of the bar. It got cross threaded either when it was being put into position or when it was being taken off. During the procedure they were not able to disconnect it; even using pliers. After the procedure, the bolt was able to be removed with some struggle. It was noted when the device had been assembled before the procedure, it was a little stiff, but it went together nicely. The procedure was successfully completed with no delay reported. This report is for a spindle nut. This is report 2 of 2 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSPINDLE NUT
Type of DeviceAPPARATUS, TRACTION, NON-POWERED
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kara ditty-bovard
1302 wrights lane east
west chester, PA 19380
6103142063
MDR Report Key10817952
MDR Text Key215572769
Report Number2939274-2020-05070
Device Sequence Number1
Product Code HST
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial,Followup
Report Date 10/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/10/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number394.42
Device Catalogue Number394.42
Device Lot Number4L61264
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/04/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received12/01/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

Patient Treatment Data
Date Received: 11/10/2020 Patient Sequence Number: 1
Treatment
14.0MM THREADED SPINDLE 480MM
-
-