• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. IV3000 1 HAND 6X7CM CTN 100; TAPE AND BANDAGE, ADHESIVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. IV3000 1 HAND 6X7CM CTN 100; TAPE AND BANDAGE, ADHESIVE Back to Search Results
Catalog Number 4007
Device Problem Material Integrity Problem (2978)
Patient Problems No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/17/2020
Event Type  malfunction  
Event Description
It was reported that when the pouch was opened, it was confirmed the film had been creased, so it was not used for treatment.Please confirm the attached photo of the sample.A backup was available.No delay was reported.No case involved.
 
Manufacturer Narrative
H10: h6: we have now concluded our investigation for the complaint received.A review of the associated batch manufacturing record did not identify any issue at the point of manufacture which may have caused or contributed to the issues highlighted by the complainant.In addition it can be confirmed that all finished product specification testing was satisfied at the point of release.A complaint history review was carried out using the lot and the part number provided, there have been further complaints reported with this failure mode in the past three years.It was reported that the dressings were used for treatment and creasing was found on the film.The returned samples were visually and functionally evaluated which confirmed this issue.There are checks in place to prevent creasing of the film.If creasing is identified during in-process checks, it is tabbed and recorded in the fault log.On this occasion the operator appears to have missed the issue.Minor creasing can sometimes occur in the film during the adhesive application process at the start or end of the roll.Again visual checks are in place to identify this.The root cause was a raw material issue and a relationship was confirmed between the device and event.As the occurrence of this failure mode is within acceptable range, no further actions by smith and nephew are deemed necessary at this stage.However, we will continue to monitor for any adverse trends relating to this product range.Smith and nephew acknowledge customer concern and are continually investigating ways to develop and improve our products.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IV3000 1 HAND 6X7CM CTN 100
Type of Device
TAPE AND BANDAGE, ADHESIVE
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull HU3 2 BN
UK  HU3 2BN
MDR Report Key10821135
MDR Text Key215880328
Report Number8043484-2020-03839
Device Sequence Number1
Product Code KGX
UDI-Device Identifier05000223417383
UDI-Public05000223417383
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup
Report Date 01/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/11/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number4007
Device Lot Number201945
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/04/2020
Date Manufacturer Received01/04/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-