Catalog Number IGTCFS-65-1-FEM-CELECT
Device Problem Structural Problem (2506)
Patient Problems Internal Organ Perforation (1987); Perforation of Vessels (2135); Great Vessel Perforation (2152); Bowel Perforation (2668)
Event Type  Injury  
Manufacturer Narrative
Manufacturer ref# (b)(4).Catalog# is unknown but referred to as cook celect filter.Occupation: non-healthcare professional.Summary of investigational findings: the following allegations have been investigated: perforation of inferior vena cava (ivc) and small bowel.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.Cook medical will continue to monitor for similar events.
 
Event Description
Description of event according to initial reporter: it is alleged that "on or about (b)(6) 2011, [pt] was implanted with a cook celect ivc filter.The cook filter placed in [pt] was stated to be appropriate for use as a permanent filter or a retrievable filter.On (b)(6) 2019, [pt] underwent a computerized tomography scan ("ct scan") of her abdomen.On or about (b)(6) 2019, [pt] was informed that the ct scan showed that multiple struts are perforating through her ivc and into her small bowel." patient outcome: it is alleged that "[pt] faces numerous health risks, including the risks of death.Because the filter cannot be removed, [pt] will require ongoing medical care and monitoring for the rest of her life.".
 
Manufacturer Narrative
Additional information: investigation: investigation is reopened due to additional information provided.Per quality engineering review, the additional information provided for this complaint does not change the previous investigation conclusion.Therefore, no new investigation activities will be conducted at this time.The catalog # and lot # are unknown, but the filter is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
Patient allegedly received the celect filter via the left femoral vein.Report from ct (computed tomography): "ivc: presence of an ivc filter below the level of the renal veins.Multiple struts of the ivc filter projects outside the boundaries of the ivc.Few struts extend into the adjacent crossing small bowel loops.Evaluation for ivc patency is somewhat limited due to more arterial phase of acquisition.No focal distention or obvious filling defect to suggest ivc thrombosis.".
 
Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Investigation: investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.The following allegations have been investigated: organ/vena cava (vc) perforation.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma.20 devices in lot.No other complaints have been reported against the lot.The associated work order was reviewed.No related/relevant notes were documented.The device is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOK CELECT NAVALIGN FEMORAL VENA CAVA FILTER SET
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
DA  4632
Manufacturer (Section G)
WILLIAM COOK EUROPE
sandet 6, dk-4632
bjaeverskov
Manufacturer Contact
henriette s. christiansen
sandet 6, dk-4632
bjaeverskov 
56868686
MDR Report Key10822075
MDR Text Key217336625
Report Number3002808486-2020-01031
Device Sequence Number1
Product Code DTK
UDI-Device Identifier10827002529141
UDI-Public(01)10827002529141(17)131108(10)E2668299
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K090140
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 03/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/11/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/08/2013
Device Catalogue NumberIGTCFS-65-1-FEM-CELECT
Device Lot NumberE2668299
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received02/28/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/01/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
Patient SexFemale
-
-