MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION 2000 ML TPN BAG; SET, I.V. FLUID TRANSFER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION 2000 ML TPN BAG; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H938740
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
This event occurred between (b)(6) 2020.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that ten (10) 2000ml eva (ethyl vinyl acetate) tpn (total parenteral nutrition) bags had issues of failed or bad clamps.It was further reported that shards of plastic break off when closing the clamps.This was identified during setup prior to patient use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
The lot was manufactured from june 05, 2020 - june 06, 2020.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The devices were not received for evaluation; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2000 ML TPN BAG
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key10824330
MDR Text Key215981834
Report Number1416980-2020-07024
Device Sequence Number1
Product Code LHI
UDI-Device Identifier00085412477244
UDI-Public(01)00085412477244
Combination Product (y/n)Y
PMA/PMN Number
K900585
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 12/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/11/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date05/31/2023
Device Catalogue NumberH938740
Device Lot Number60244072
Was Device Available for Evaluation? No
Date Manufacturer Received12/16/2020
Is This a Reprocessed and Reused Single-Use Device? No