• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION REVACLEAR 300; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION REVACLEAR 300; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 114748
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Patient Involvement (2645)
Event Date 10/17/2020
Event Type  malfunction  
Manufacturer Narrative
Phone number: (b)(6).Facility name: (b)(6) hospital.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a particulate matter was observed inside the cap of a revaclear 300 dialyzer before use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
H10: the actual device was not available; however, two photographs of the sample were provided for evaluation.During the visual inspection a point (particle) in the header area was visible.However, due to the poor image quality, it is not visible where (inside, outside, is the particle injected or loose ) the particle is.Only the product label is visible on the second photo.The cause could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REVACLEAR 300
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key10825250
MDR Text Key215983085
Report Number9611369-2020-00158
Device Sequence Number1
Product Code KDI
UDI-Device Identifier07332414114947
UDI-Public(01)07332414114947
Combination Product (y/n)Y
PMA/PMN Number
NI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 12/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/11/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2023
Device Catalogue Number114748
Device Lot Number0-9608-H-01
Was Device Available for Evaluation? No
Date Manufacturer Received12/01/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-