MAUDE Adverse Event Report: SMITH & NEPHEW, INC. MOTOR DRIVE UNIT HAND CNTRL PWRMX EL; SAW, POWERED, AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. MOTOR DRIVE UNIT HAND CNTRL PWRMX EL; SAW, POWERED, AND ACCESSORIES Back to Search Results
Model Number 72200616
Device Problem Mechanical Jam (2983)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/12/2020
Event Type  malfunction  
Event Description
It was reported that, during an acl repair surgery, the "mdu powermax elite shaver" stalled.The procedure was successfully completed without delay using a back-up device.No patient injuries or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
Internal complaint reference: (b)(4).The reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.A complaint history review concluded this was a repeat issue.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.
 
Manufacturer Narrative
H10: h2: additional information: h6: health effect - clinical and impact code.H3, h6: the reported device was received for evaluation.It was determined the device contributed to the reported event.A visual inspection was performed on the exterior of product and no physical damage was observed.Product failed functional testing with blade stall error and overheating.Cause of overheating and errors is a corroded motor/gearbox.The gearbox was found to be jammed.The complaint was confirmed and the root cause has been determined to be corrosion of the motor and gearbox assembly.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review concluded this was a repeat event.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The complaint was confirmed, and the root cause was associated with corrosion in the gearbox.Factors which can contribute to overheating include a blade stall condition that will result in increased current draw from the control unit which will heat the motor and hand piece housing.This can be the result of gearbox corrosion caused by cleaning and sterilization methods and the chemicals involved.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.Several corrective actions were taken to reduce the occurrence of this failure.No further actions are required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOTOR DRIVE UNIT HAND CNTRL PWRMX EL
Type of Device
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key10825446
MDR Text Key215938691
Report Number1643264-2020-01786
Device Sequence Number1
Product Code HAB
UDI-Device Identifier03596010555472
UDI-Public03596010555472
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 01/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/11/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72200616
Device Catalogue Number72200616
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/27/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-