• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. RENASYS GO; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. RENASYS GO; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP Back to Search Results
Model Number 66801496
Device Problem Complete Blockage (1094)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/21/2020
Event Type  malfunction  
Event Description
It was reported that during testing, unit stops when trying to complete blockage test.No patient impact reported.
 
Manufacturer Narrative
H3, h6: the device, intended for use in treatment, was returned for evaluation.Visual inspection of the returned device noted a cracked lower and upper casing.Functional evaluation revealed that the battery failed to charge and there was a recharge, ex low battery error message.The device shuts down due to depleted battery power.Establishing a relationship between the device and the reported event.The root cause was identified as a depleted battery.The manufacturing records show no evidence that the product did not meet the specification at the time of manufacture.A complaint history review found further instances of the reported event.This investigation is now complete with no further action deemed necessary.Smith + nephew will continue to monitor for any adverse trends relating to this product range.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RENASYS GO
Type of Device
NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
MDR Report Key10826010
MDR Text Key215881488
Report Number8043484-2020-03861
Device Sequence Number1
Product Code OMP
UDI-Device Identifier00040565126944
UDI-Public00040565126944
Combination Product (y/n)N
PMA/PMN Number
K152163
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Type of Report Initial,Followup
Report Date 07/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/11/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number66801496
Device Catalogue Number66801496
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/09/2021
Date Manufacturer Received07/02/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-