Model Number PM3562 |
Device Problems
Incorrect Measurement (1383); Data Problem (3196)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/29/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined. .
|
|
Event Description
|
During follow-up, a diagnostic anomaly when measuring capture and impedance occured.No intervention was performed, the device remains implanted and will continue to be monitored.The patient was in stable condition.
|
|
Manufacturer Narrative
|
A software investigation was performed and the reported event of a diagnostic anomaly resulting in unavailable impedance and right ventricular cap confirm threshold measurements was confirmed.Based on the review of the information, the measurements were continuously being delayed.The root cause of the event is due to the simultaneous occurrence of device measurements.This can be resolved by changing the timing of these two measurements so they do not occur at the same time, cancelling each other out.
|
|