Model Number B1100-060
Device Problem Device Damaged by Another Device (2915)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/23/2020
Event Type  malfunction  
Manufacturer Narrative
The device is not returning for evaluation. Investigation is not yet complete. A follow-up report will be submitted with all additional relevant information. The additional device referenced is being filed under a separate medwatch report#.
 
Event Description
It was reported that during a transjugular intrahepatic portosystemic shunt procedure, a 10. 0x100mm absolute pro stent was deployed on a bend. After dilatation with a 10. 0x60mm armada 35 balloon dilatation catheter, it was observed that more than one stent strut fractured in the middle of the stent but the stent remained as one. The fractured stent remains in the patient. There were no adverse patient effects and no clinically significant delay in the procedure. No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameARMADA 35 PTA CATHETER
Type of DevicePERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key10838183
MDR Text Key216338549
Report Number2024168-2020-09515
Device Sequence Number1
Product Code LIT
Combination Product (y/n)N
Reporter Country CodePM
PMA/PMN Number
K111899
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation
Type of Report Initial,Followup
Report Date 12/02/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received11/13/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date03/31/2022
Device Model NumberB1100-060
Device Catalogue NumberB1100-060
Device Lot Number81004G1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received11/17/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/04/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-