• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - PUMP; VENTRICULAR (ASSIST) BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - PUMP; VENTRICULAR (ASSIST) BYPASS Back to Search Results
Model Number 1103
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Dyspnea (1816); Respiratory Distress (2045); Respiratory Failure (2484)
Event Date 10/14/2020
Event Type  Injury  
Manufacturer Narrative
Investigation of this event is pending and a supplemental report will be sent upon its completion.This information was received from the vad destination therapy post approval study.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the ventricular assist device (vad) patient was reintubated for respiratory distress.The patient underwent at racheostomy procedure.The vad remains in use.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
A supplemental report is being submitted for additional information, patient codes and health impact codes were updated and device evaluation.Product event summary: the ventricular assist device (vad) was not returned for evaluation.This complaint is associated with a clinical adverse event.Information provided by the site indicated that the patient was reintubated for respiratory distress and underwent a tracheostomy procedure.Based on the available information, there is no evidence to indicate that a device malfunction or performance issue caused or contributed to the reported event.Per the instructions for use, respiratory dysfunction is a known potential complication associated with the implantation of a vad.Based on review of past adverse events for this patient, it was noted that the patient had a history of respiratory dysfunction.Possible clinical factors that may have contributed to this event include the patient¿s pre-existing history and related comorbidities, the progression of their underlying disease, issues related to the therapeutic use of anticoagulant and antiplatelet medications and the patient's complex post-operative course.There are possible patient, pharmacological and procedural factors that may have contributed to this event.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE VENTRICULAR ASSIST SYSTEM - PUMP
Type of Device
VENTRICULAR (ASSIST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
MDR Report Key10840014
MDR Text Key216315408
Report Number3007042319-2020-07313
Device Sequence Number1
Product Code DSQ
UDI-Device Identifier00888707003261
UDI-Public00888707003261
Combination Product (y/n)N
PMA/PMN Number
P100047
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,study
Type of Report Initial,Followup
Report Date 01/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/13/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date02/28/2022
Device Model Number1103
Device Catalogue Number1103
Was Device Available for Evaluation? No
Date Manufacturer Received01/07/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Required Intervention;
Patient Weight90
-
-