Model Number TSH
Device Problems Incorrect, Inadequate or Imprecise Resultor Readings (1535); Non Reproducible Results (4029)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/21/2020
Event Type  malfunction  
Manufacturer Narrative
This event occurred in (b)(6). The investigation is ongoing.
 
Event Description
The initial reporter received questionable elecsys tsh assay, elecsys ft4 iii assay, and elecsys ft3 iii results for one patient tested on a cobas 8000 e 602 module serial number (b)(4). The customer reported out the results to a physician who asked for the re-measurement of the sample. The patient¿s sample was provided for an investigation and was tested on a cobas 8000 e 602 module. Also, the patient's sample was outsourced and tested on an abbott architect analyzer and siemens centaur analyzer. Refer to the attachment on the medwatch for all patient data. This medwatch is for tsh. Refer to the medwatch with a1 patient identifier (b)(4) for the ft4 assay and (b)(4) for the ft3 assay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameELECSYS TSH ASSAY
Type of DeviceRADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM 68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key10844305
MDR Text Key227502481
Report Number1823260-2020-02909
Device Sequence Number1
Product Code JLW
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K961491
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/29/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received11/16/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberTSH
Device Catalogue Number11731459122
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received10/26/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Invalid Data

-
-