The user facility reported to terumo cardiovascular that during cardiopulmonary bypass, there was a plasma leakage.
As per the sales associate, there was a negligible amount of blood loss and the customer resolve the issue by increasing the sweep to 2 to 4.
As per the clinical specialist, the director of clinical specialist called the perfusionist on (b)(6) 2020.
The perfusionist who started the case reported no issues during the hand off discussion.
The plasma leak took place at the ending stages of the 5 hour and 13 minute procedure for a cabg x 4 aortic valve replacement procedure on a 72 y/o male.
The perfusionist noted that this patient had been hospitalized 2 days prior to the surgery which is unusual, but he didn't know why.
He didn't feel that the patient exhibited any inflammatory tendencies while on bypass although he did note that he needed to run his sweep gas higher than what he has experienced in the past, 3.
5 to 4 lpm versus 2lpm typically.
It was not at all unmanageable and po2 was not affected at all, so during the conversation he concluded that it was most likely patient to patient variability.
Another thing he noted was that the scavenge line on their boehringer vavd controller seemed to have more condensate than normal.
But again, slower surgeon, potentially longer rewarm time.
The patient was large - 102 kilos.
They had to run a higher sweep than what they consider normal - 4 lpm versus typically running 2 lpm.
Gas exchange performance was not a problem.
There was a negligible amount of blood loss; the product was not changed out; procedure completed successfully.
|