ABBOTT VASCULAR TREK CORONARY DILATATION CATHETER Back to Search Results
Model Number 1012280-15
Device Problems Deflation Problem (1149); Off-Label Use (1494); Material Separation (1562)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/26/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4). The device was received. Investigation is not yet complete. A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported that the procedure was to treat a left pulmonary artery. A 5. 0x15mm trek balloon was advanced and inflated at the first inflation at 8 atmospheres. The balloon would not deflate after attempting multiple times with different size syringes. Therefore, the balloon was intentionally inflated more to rupture. Once the balloon ruptured, during removal the shaft separated into two pieces and a snare was used to retrieved the separated portion. No other device was used to complete the procedure. There were no adverse patient sequela and no clinically significant delay. No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameTREK CORONARY DILATATION CATHETER
Type of DeviceCORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key10853234
MDR Text Key216862735
Report Number2024168-2020-09608
Device Sequence Number1
Product Code LOX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K103110
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial,Followup
Report Date 12/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/17/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date07/31/2022
Device Model Number1012280-15
Device Catalogue Number1012280-15
Device Lot Number90807G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/06/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received12/02/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/07/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 11/17/2020 Patient Sequence Number: 1
-
-