Model Number 284002
Device Problem Increase in Pressure (1491)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Udi: (b)(4). Investigation summary: the device was received and evaluated at the service center. The reported complaint that the fms vue pump will not power up was confirmed. The main pcb was replaced to correct the power up issue. Also replaced electrovanne due to excessive pressure. Also unit rebooted during test, replaced power supply and performed hipot test. The device was found to be working according to specifications. However, given the information provided, we cannot discern a definitive root cause of the defective parts. The service history has been reviewed in lieu of the device history record for this device since it was previously serviced. The device was last serviced on 03/12/2020 and passed all functional testing before being returned to the customer. At this point in time, no corrective action is required, and no further action is warranted. Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Event Description
It was reported that preoperatively to an unknown procedure on an unknown date, it was observed that the pump device would not power up. During in-house engineering evaluation, it was determined that the electrovanne had excessive pressure. It was not reported if there was a delay in the surgical procedure. It was not reported if a spare device was available for use. This event did not occur during surgery. There was no patient involvement. There were no reports of injuries, medical intervention or prolonged hospitalization. All available information has been disclosed. If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameFMS VUE PUMP
Type of DeviceDISTENSION UNIT, FLUID, ARTHROSCOPIC
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6013142063
MDR Report Key10856204
MDR Text Key217331449
Report Number1221934-2020-03524
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705020591
UDI-Public10886705020591
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K130169
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/09/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received11/17/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number284002
Device Catalogue Number284002
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/04/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received11/09/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-