Model Number 66021496
Device Problem No Apparent Adverse Event (3189)
Patient Problem No Patient Involvement (2645)
Event Date 10/01/2020
Event Type  malfunction  
Event Description
It was reported that during an inspection the keypad freezed up.A backup was available.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H3, h6: the device, intended for use in treatment, was returned for evaluation.A visual inspection reported cracked lower and upper casing.The functional evaluation found no faults with unlocking the screen, by pressing the unlock button for 3 seconds , establishing no relationship between the device and the reported event.A probable cause is user error a review of the manufacturing records found that there was no evidence that the product didn¿t meet specifications at the time of manufacture.A complaint history review found further instances of the reported events.This investigation is now complete with no further action deemed necessary.Smith + nephew will continue to monitor for any adverse trends relating to this product range.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RENASYS GO
Type of Device
NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
MDR Report Key10857647
MDR Text Key216820467
Report Number8043484-2020-03895
Device Sequence Number1
Product Code OMP
UDI-Device Identifier05000223491000
UDI-Public05000223491000
Combination Product (y/n)N
PMA/PMN Number
K152163
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup
Report Date 07/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/17/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number66021496
Device Catalogue Number66021496
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/27/2021
Date Manufacturer Received07/08/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-