MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-722WWL; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-722WWL
Device Problem Obstruction of Flow (2423)
Patient Problems Hyperglycemia (1905); Scar Tissue (2060)
Event Date 09/12/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
The customer reported via phone call that customer were experiencing high blood glucose.Blood glucose level was 400 mg/dl.Customer stated insulin pump had insulin flow blocked alarm.Customer stated they had a lot of scar tissue.Customer declined troubleshooting for high blood glucose.The insulin pump will be returned for analysis.
 
Manufacturer Narrative
Device alarmed motor error during rewind due to motor encoder signal out of phase.Unable to perform the displacement test, rewind, basic occlusion test, occlusion test, prime or a33 test and excessive no delivery test due to motor error alarms.Device received with cracked belt clip slot, and cracked reservoir tube lip.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-722WWL
Type of Device
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key10857997
MDR Text Key216820090
Report Number2032227-2020-206154
Device Sequence Number1
Product Code OYC
UDI-Device Identifier000000613994401991
UDI-Public(01)000000613994401991
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 12/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/17/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberMMT-722WWL
Device Catalogue NumberMMT-722WWL
Device Lot NumberB2722WWLJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/24/2020
Date Manufacturer Received12/08/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
FRN-MMT-332A-RSVR, UNOMED INF SET; FRN-MMT-332A-RSVR, UNOMED INF SET
Patient Outcome(s) Other;
Patient Age45 YR
Patient Weight165
-
-