Product analysis: the spiderfx device was returned within its shelf box, sterile pouch and transportation hoop.No ancillary devices, cine, images or procedure notes were returned for evaluation.The device was removed from the hoop for analysis.Visual inspection of the catheter shaft reveal no physical damage (kink, twist).No abnormality was noted on the catheter distal tip and capture wire.The filter was observed within a clear segment.A 0.014in guidewire from analysis lab was loaded successfully through the green dual delivery catheter.The guidewire would advance and exit the rx exit port with no resistance noted.When the capture wire was attempted to advance the filter out of the clear segment , it was met with resistance and the filter would not advance.The catheter was soaked in for 3hrs in ultrasonic bath.Post soaking, it was observed the flex connector was detached within the filter basket.The damage observe d is consistent with resistance encountered during advancement.If information is provided in the future, a supplemental report will be issued.
|