RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problems
Electrical /Electronic Property Problem (1198); Power Problem (3010)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2020.Date of report: 18nov2020.
|
|
Event Description
|
The biomed is reporting the v60 is failing the electrical safety grounding test at the oxygen inlet.The remote service engineer (rse) advised the biomed to remove excess loctite from the oxygen inlet bracket retaining screws.The biomed is reporting they have attempted to remove the loctite but have not cleaned the oxygen inlet fitting retainer bracket.The device did not have patient involvement at the time the issue was discovered, therefore, there was no patient or user harm reported.
|
|
Manufacturer Narrative
|
H10: the customer advised philips to close the case.The alleged device malfunction could not be confirmed as no repair information was provided.No product was returned for failure investigation.The complaint will be closed.If additional information is received at a later date, the complaint will be reopened, and a supplemental report will be submitted.
|
|