Model Number MI1200 SYNCHRONY |
Device Problems
Impedance Problem (2950); Output Problem (3005)
|
Patient Problem
Failure of Implant (1924)
|
Event Date 11/16/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device has not been explanted.If it should be explanted, it is to be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
|
Event Description
|
The user's hearing performance with the device is affected.There has been a progressive number of channels affected by high impedance.
|
|
Event Description
|
The user's hearing performance with the device is affected.There has been a progressive number of channels affected by high impedance.Medical intervention is considered but no date has been scheduled yet.
|
|
Manufacturer Narrative
|
Additional information: according to the currently available information, damage to the active electrode likely caused by minute device mobility is suspected.However to determine an exact root cause a device investigation of the explanted device is necessary.Further medical intervention is considered but no date has been scheduled yet.
|
|
Manufacturer Narrative
|
Conclusion: damage to the active electrode which is consistent with minute device mobility was determined to have led to device failure over time due to fatigue wire breakages.The problems given in the recipient report appear to match the damage found.This is a final report.
|
|
Event Description
|
The user_s hearing performance is affected.There has been a progressive number of channels affected by high impedance.The user was reimplanted on the (b)(6) 2022.
|
|