Catalog Number 955467
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Patient Involvement (2645)
Event Date 10/17/2020
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that during priming with a prismaflex m150 set, external fluid leakage was observed.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information: h6 and h10.H10: the actual device was not available; however, a photograph of the sample was provided for evaluation.The visual inspection observed a fluid leak at the level of the upper dialysate port.The reported condition was verified.The cause is due to mechanical shock during transportation.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX M150
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key10870178
MDR Text Key217705848
Report Number8010182-2020-00256
Device Sequence Number1
Product Code KDI
Combination Product (y/n)Y
PMA/PMN Number
NI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 12/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/19/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number955467
Device Lot Number20A2604M
Was Device Available for Evaluation? No
Date Manufacturer Received11/24/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-