• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES EDWARDS ESHEATH INTRODUCER SET; PROSTHESIS, MITRAL VALVE, PERCUTANEOUSLY DELIVERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES EDWARDS ESHEATH INTRODUCER SET; PROSTHESIS, MITRAL VALVE, PERCUTANEOUSLY DELIVERED Back to Search Results
Model Number 914ESA
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Vascular System (Circulation), Impaired (2572); Unspecified Vascular Problem (4441)
Event Date 10/26/2020
Event Type  Death  
Manufacturer Narrative
Exemption number e2016006, this report summarizes 2 events of major vascular complication for the sapien 3 in the mitral position.The ¿time to event¿ (tte, in days) for this event was 2.   the device identification (di) numbers for edwards esheath introducer set is (b)(4).  according to the instructions for use (ifu), cardiovascular complications, including perforation or dissection of vessels which may require intervention, are potential adverse events associated with the transfemoral transcatheter aortic valve replacement procedure.According to literature review, and as documented in a clinical technical summary written by edwards lifesciences, vascular complications are a well-recognized complication of the transfemoral tavr procedure in this elderly population with multiple co-morbidities.Edwards has reviewed many reports, including screening data records and source documentation of vascular complications and has found that the root cause is typically related to a combination of vessel size, tortuosity and calcifications.Although the incidence is decreasing with smaller sheath/delivery system sizes and physician experience, there will continue to be cases in which vascular complications will occur.The thv physician training manuals instruct on procedural considerations for sheath insertion with regards to proper screening critical to reducing vascular complications.The training manual instructs the operator on proper sheath insertion and withdrawal techniques, including pre-dilating the vessel with the edwards dilators, as needed.It also notes that calcification may reduce lumen diameter and limit or prevent transfemoral passage of the devices.The ifu contraindicates patients with ilio-femoral vessel characteristics that would preclude safe placement of sheaths such as severe obstructive calcification or severe tortuosity.Pre-procedure screening and assessment of the femoral/iliac artery internal diameters will enable the clinician to determine if the thv can be delivered transfemorally.Assessment of location and amount of circumferential calcium will aid in determining areas of reduced vessel diameters.The operators are trained to measure minimum vessel diameter taking calcium into account.The physician training manual also lists the minimum recommended vessel size for each size device.Despite the best screening tools, a small percentage of patients will have femoral/iliac vessels that are not amenable to the trans-femoral approach or where increased resistance is encountered during insertion of devices.In many cases, the vessel minimum luminal diameter (mld) may be borderline or below the indicated size.In addition, significant calcification and/or tortuosity, not always appreciable on imaging, could be contributing factors to the event.In this case, specific procedural details are not available to determine potential contributing factors to the event, or if the event is related to an edwards device.The ifu and training manuals have been reviewed and no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.No corrective or preventative actions are required at this time.
 
Event Description
Thv/tvt registry alternative summary reporting (asr) adverse event submission for august 2020 data extract for mitral deaths for the sapien 3.August 2020 data extract includes data provided by acc for q1 2020 (january 1 ¿ march 31, 2020) .  this report summarizes 2 death events of major vascular complication for the sapien 3 for august 2020.The age for these events is (b)(6).The breakdown for gender is as follows: 2 females.
 
Manufacturer Narrative
Supplemental report to add noe statement in b5 section, and provide d5, and h6 information.
 
Event Description
This report summarize  2  death events of  major vascular complication for the sapien 3 for (b)(6) 2020.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS ESHEATH INTRODUCER SET
Type of Device
PROSTHESIS, MITRAL VALVE, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
MDR Report Key10871936
MDR Text Key217293536
Report Number2015691-2020-14505
Device Sequence Number1
Product Code NPU
Combination Product (y/n)N
PMA/PMN Number
P140031
Number of Events Reported2
Summary Report (Y/N)Y
Report Source Manufacturer
Source Type health professional,user faci
Remedial Action Other
Type of Report Initial,Followup
Report Date 10/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/19/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number914ESA
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? No
Date Manufacturer Received11/19/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Death;
Patient Age72 YR
-
-