Model Number PED-475-25 |
Device Problems
Material Deformation (2976); Activation Failure (3270)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/18/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received a report that during the pipeline release process, the distal end was bent, flattened, and could not be opened completely.After being opened by a push-pull operation, the middle section could not be opened.Replacement products were then used to complete the procedure.It was indicated that all devices were prepared as per the instructions for use (ifu), and no patient symptoms or further complications were reported as a result of this event.Post-procedure angiographic results showed slowed blood flow in the aneurysm.The patient was undergoing surgery for treatment of a saccular, unruptured aneurysm of the ophthalmic artery with a max diameter of 5.2 mm and a 5.5 mm neck diameter.It was noted the patient's vessel tortuosity was severe.
|
|
Event Description
|
Additional information received reporting that the pipeline was placed in a vessel bend when the failure to open occurred.It was unknown how many times the pipeline was resheathed.The physician also attempt to use the catheter and pipeline pushwire to coax the pipeline to open without success.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
H3: analysis of the pipeline flex (lot no.B052130) found that the pipeline flex could not be pushed forward or removed.For further examination, the catheter was cut to remove the pipeline flex delivery system.The distal and proximal dps restraints were found to be intact.The dps sleeves were found intact with no signs of damage.The distal hypotube appeared to be stretched with the ptfe shrink tubing still intact.The distal and proximal ends of the pipeline flex braid were found fully opened and moderately frayed.The middle section of the braid appeared to be fully opened with no damage.Bends were found at 12.0 cm to 38.0 cm from the proximal end.No defects were found with the tip coil, distal marker, re-sheathing marker, re-sheathing pad or with the proximal bumper.The total and usable lengths of the marksman catheter were measured to be within specifications.The catheter tip and marker were examined; no damages were found.The catheter body appeared to be accordioned at 25.0 cm to 31.0 cm from the distal tip.No flash or voids molded were observed in the hub.The catheter was then tested by running an in-house 0.0265¿ mandrel through catheter tip and hub.The mandrel successfully passed through the catheter hub and tip; however, resistance was observed at the damaged locations.No other anomalies were observed.Based on the returned devices, the pipeline flex was not confirmed to have "failure to open" issue as the middle section of the pipeline flex braid was found fully opened with no damage.In addition, the distal and proximal ends of the pipeline flex were fully opened and moderately frayed.The damage to the braid on the ends of the pipeline flex is likely the results of the physician re-sheathing the device more than recommended two times.Furthermore, the pipeline flex delivery system was found stuck inside the distal segment of the marksman catheter.From the damages seen on the catheter (accordioning), pusher (bending), pipeline flex braid (fraying) and hypotube (stretching); it appears there was high force used.It is likely these damages occurred when the customer attempted to advance the pipeline flex through the marksman catheter against the resistance.It is likely that the severe vessel tortuosity and lack of continuous flush with heparinized saline may have contributed to the failure to open and resistance issues.Per our instructions for use (ifu), the user should: ¿begin to deliver the device using a combination of unsheathing the pipeline flex embolization device and pushing delivery wire simultaneously.After the distal end of the pipeline flex has successfully expanded, deploy the remainder of the device.The system is designed to allow for 2 full cycles of re-sheathing of the pipeline flex embolization device.Re-sheathing the pipeline flex embolization device more than 2 full cycles may cause damage to the distal or proximal ends of the braid.Discontinue delivery of the device if high force or excessive friction is encountered during delivery.Identify the cause of the resistance and remove device and microcatheter simultaneously.Advancement of the ped against resistance may result in device damage or patient injury.Never advance or withdraw an intraluminal device against resistance until the cause of resistance is determined by fluoroscopy.If the cause cannot be determined, withdraw the catheter.Movement of the micro catheter against resistance may result in damage to the micro catheter, or the vessel.Do not use in patients in whom the angiography demonstrates the anatomy is not appropriate for endovascular treatment, due to conditions such as severe intracranial vessel tortuosity or stenosis.¿ h6: method code updated to b01.Result code updated to c070601.Conclusion code updated to d1103.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
No additional information received.
|
|