• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US UNKNOWN HIP ACETABULAR CUP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US UNKNOWN HIP ACETABULAR CUP Back to Search Results
Catalog Number UNK HIP ACETABULAR CUP
Device Problems Device Dislodged or Dislocated (2923); Naturally Worn (2988)
Patient Problems Foreign Body Reaction (1868); Pain (1994); Osteolysis (2377); No Code Available (3191)
Event Date 04/28/2020
Event Type  Injury  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
The literature article entitled, "protrusion of a ceramic femoral head through the acetabular metallic cup in total-hip arthroplasty a case report" written by ding zhao, md, de-bao zhang, md, dong-feng han, md, and gui-shan gu, md published by medicine accepted by publisher on 28 april 2020 was reviewed.The article reports on one case of a (b)(6) male who received tha in 2010 in right hip with titanium 50 mm acetabular cup, poly liner, size 8 trilock bps stem, 28 mm femoral ceramic head (all implants from depuy) underwent a revision in 2018.Article notes he did not perform follow ups and often participated in strenuous sports activities including mountain climbing and long distance running.He reported pain since 2017 in right groin.By june 2018 the pain increased and radiographs revealed internal superior subluxation of the head within the cup.Intraoperative findings were a disassociated liner which lead to the liner being impinged with femoral neck.The liner had portions completely worn through leading the ceramic head in direct contact with the femoral cup leading to the cup being worn through as well.Intraoperative findings also include black pigmentation of metallic debris within osteolytic lesions and around the joint capsule and acetabulum.All components were removed including the stem and replaced with new depuy implants.No further complications.The article provides photographic images for illustrative purposes.Depuy products: poly liner, titanium cup, trilock stem, ceramic head.Adverse events: disassociation of liner from cup (root cause which lead to poly liner wear, metal wear and metallic debris) - treated by revision of all components.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.Review of the attached images and x-rays confirmed the reported complaint.The root cause could not be determined.The information received will be retained for potential series investigations if triggered by trend analysis or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the lot code was provided, a manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN HIP ACETABULAR CUP
Type of Device
HIP ACETABULAR CUP
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10879595
MDR Text Key217640923
Report Number1818910-2020-25179
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,l
Type of Report Initial,Followup
Report Date 11/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/20/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNK HIP ACETABULAR CUP
Was Device Available for Evaluation? No
Date Manufacturer Received02/08/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age61 YR
-
-