• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. DII CONTROLLER; ARTHROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. DII CONTROLLER; ARTHROSCOPE Back to Search Results
Model Number 72200873
Device Problem Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/28/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that during an arthroscopy surgery the controller is not accepting foot pedal or the hand control pieces.They were not being recognized.The procedure was successfully completed without a significant delay using a back-up device.No patient injury or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
The reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of manufacturing records found the device has been in service for over 5 years, thus any malfunction presented at this point in time would not be related to the manufacture of the product.A complaint history review concluded this was a repeat issue.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.
 
Manufacturer Narrative
The reported device, used in treatment, was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Complaint of mdu and footswitch port malfunctions could not be reproduced.Product passed functional testing with no faults or errors.Both mdu ports and the footswitch port perform as expected.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.The device has been in service for over 5 years, thus any malfunction presented at this point in time would not be related to the manufacture of the product.A complaint history review concluded this was a repeat issue.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DII CONTROLLER
Type of Device
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
MDR Report Key10881694
MDR Text Key217697039
Report Number1643264-2020-01902
Device Sequence Number1
Product Code HRX
UDI-Device Identifier03596010607409
UDI-Public03596010607409
Combination Product (y/n)N
PMA/PMN Number
K062849
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Type of Report Initial,Followup,Followup
Report Date 03/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/22/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72200873
Device Catalogue Number72200873
Was Device Available for Evaluation? No
Date Manufacturer Received03/08/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-