• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHYSIO-CONTROL, INC. - 3015876 LIFEPAK 20E DEFIBRILLATOR/MONITOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC. - 3015876 LIFEPAK 20E DEFIBRILLATOR/MONITOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number 20E
Device Problem Device Sensing Problem (2917)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/25/2020
Event Type  malfunction  
Manufacturer Narrative
Physio-control contacted the customer to request additional information on the patient.No response has been received from the customer.Patient fields in which information was not provided were intentionally left blank.Physio-control evaluated the customer's device and was unable to duplicate the reported issue.After observing proper device operation through functional and performance testing, the device was returned to the customer for use.The cause of the reported issue could not be determined.
 
Event Description
The customer contacted physio-control to report that they were using this device on a patient and the device was not showing the patient's ecg rhythm through the paddles lead.As a result, defibrillation therapy may have been delayed or unavailable, if needed.There were no reports of any adverse effects to the patient as a result of the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK 20E DEFIBRILLATOR/MONITOR
Type of Device
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
todd bandy
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key10886590
MDR Text Key217971115
Report Number0003015876-2020-01762
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00883873852804
UDI-Public00883873852804
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P160026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 11/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/23/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number20E
Device Catalogue Number99507-000132
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/09/2020
Date Manufacturer Received10/28/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/10/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-