• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION THERANOVA 400 DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION THERANOVA 400 DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 115578
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Low Blood Pressure/ Hypotension (1914); Reaction (2414)
Event Date 10/30/2020
Event Type  Injury  
Manufacturer Narrative
(b)(6). Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that after ¿startup¿ of treatment with a theranova 400 dialyzer, the patient experienced blood pressure drop, blushing and feeling of unwellness. Treatment was discontinued. Treatment for the events was not reported. At the time of this report, the patient outcome was not reported. No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameTHERANOVA 400
Type of DeviceDIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - HECHINGEN BADEN DIALYZER
holger - crafoord - strasse 26
hechingen D-723 79
GM D-72379
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key10890976
MDR Text Key217956836
Report Number9611369-2020-00170
Device Sequence Number1
Product Code KDI
Combination Product (y/n)Y
Reporter Country CodeNO
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial,Followup
Report Date 12/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/24/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator
Device Catalogue Number115578
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received12/15/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

Patient Treatment Data
Date Received: 11/24/2020 Patient Sequence Number: 1
-
-