Reportedly, the dispatched fse could narrow down the root cause to a problem with the ventilator motor and replaced it.The log files were checked by the manufacturer and, the reported issue as well as the validity of the on-site analysis can be confirmed.The log file indicates that the procedure in question went stable and unremarkable for more than two hours until the device detected a ventilator piston position error.The device alarmed and shut down automatic ventilation.The device was used for 4 more minutes in manual ventilation before it was switched to standby.The device is equipped with a piston ventilator which is driven by a stepper motor.The number of rotations forth and back is proportional to the piston hub and to the volume that will be shifted (tidal volume).The motor speed is being monitored continuously; speed fluctuations result in a deviation between measured and expected piston position.To prevent from damages to the ventilator unit the system is designed to shut down automatic ventilation and to alert the user to this condition by means of a corresponding alarm.Manual ventilation and the monitoring functions remain available to the full extent.No patient consequences have reportedly occurred.Dräger finally concludes that the device behaved as specified upon the malfunction of a single component which is a wear-and-tear part and has lasted longer than the expected life time.The repair exchange has fully solved the device problem.The number of similar cases, related to the same root cause, is within the expected range of the respective risk assessment and thus accepted.
|