Model Number 284004
Device Problems Infusion or Flow Problem (2964); Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/06/2020
Event Type  malfunction  
Manufacturer Narrative
Udi: (b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event Description
It was reported by the affiliate in (b)(6) that during an unknown procedure on (b)(6) 2020, it was observed that the pump/shaver device did not suck; and was working on and off.The same device was used to complete the procedure with an unspecified delay.There were no adverse patient consequences reported.No additional information was provided.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary: the complaint device was received at the service center and evaluated.It was reported that ¿the device didn't suck and has been working in on and off during an unknown procedure.¿.Per service manual operational and diagnostic, no defect was found with the device, therefore this complaint cannot be confirmed.As the reported problem was not confirmed and no defects identified, a root cause for the issue that was experienced by the user cannot be determined.A manufacturing record evaluation was performed for the finished device [k09a70689] number, and no non-conformances related to the reported complaint condition were identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D4: the serial number was inadvertently left blank on the initial report; and has been updated to reflect the correct information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FMS VUE PUMP-SHAVER BOX
Type of Device
DISTENSION UNIT, FLUID, ARTHROSCOPIC
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin blanc 38
le locle CH-24 00
SZ  CH-2400
MDR Report Key10896856
MDR Text Key218489672
Report Number1221934-2020-03592
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705027934
UDI-Public10886705027934
Combination Product (y/n)N
PMA/PMN Number
K171237
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 11/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/24/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number284004
Device Catalogue Number284004
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/27/2021
Date Manufacturer Received02/04/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-