DATASCOPE CORP. - MAHWAH CS100; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Catalog Number 0998-UC-3013-53
Device Problem Display or Visual Feedback Problem (1184)
Patient Problems No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/04/2020
Event Type  malfunction  
Manufacturer Narrative
A getinge field service engineer (fse) was dispatched to evaluate this unit.The fse found the display has physical damage and needs to be replaced.Additional information has been requested, and we will report accordingly if it becomes available.
 
Event Description
It was reported that the cs100 intra-aortic balloon pump (iabp) display had blackspots, and horizontal lines on it.It is unknown the circumstances under which the event occurred.It is also unknown if there was patient involvement.However, there was no adverse event reported.
 
Event Description
It was reported that the cs100 intra-aortic balloon pump (iabp) display had blackspots and horizontal lines on it.It is unknown the circumstances under which the event occurred.It is also unknown if there was patient involvement.However, there was no adverse event reported.
 
Manufacturer Narrative
A getinge field service engineer (fse) submitted a quote; however, it is unknown if it was approved, or if it will be approved.If any pertinent information is received in the future, a supplemental report will be submitted.
 
Manufacturer Narrative
Analysis of production: (3331) the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported serial number.Based on the dhr review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis: (4109) the review of the historical data indicates that no other similar complaint was reported for the same serial number and reported failure mode.Trend analysis: (4110) the overall 24 month product complaint trend data for the period (dec 2019 through nov 2020) was reviewed.There were no triggers identified for the review period.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS100
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
MDR Report Key10899923
MDR Text Key218249240
Report Number2249723-2020-02017
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107745
UDI-Public10607567107745
Combination Product (y/n)N
PMA/PMN Number
K031636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 07/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/25/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number0998-UC-3013-53
Was Device Available for Evaluation? Yes
Date Manufacturer Received07/30/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-