Model Number T-601 |
Device Problem
Insufficient Information (3190)
|
Patient Problems
Irritation (1941); Itching Sensation (1943); Rash (2033); Burning Sensation (2146)
|
Event Type
Injury
|
Manufacturer Narrative
|
Retained samples of the concerned lots (191106-0225, 191015-0222, 191029-0155, 200121-0155, 200124-0157 and 200131-0251) of model sbt601 and model sbw601 have been inspected visually.Mechanical tests were performed on 3 retained samples.All tested samples were found to perform within limits.No faults could be detected.No information on the patient, skin type, state of skin, whether any medication was being taken, which might have a skin weakening effect, duration of use and details of the use was provided.We have requested further information several times but none was made available.The initial reporter stated "we continue to encourage the customer to provide this information.However we can't force them to do so." and in a later communication "we have spent years trying to get this provider to add additional questions to their patient complain process (forms) with no luck.I'm not sure there is much more i can do at this time." we therefore consider the investigation closed.
|
|
Event Description
|
On november 11th, 2020, we have been informed about 10 incidents with ecg electrodes.Monitoring ecg electrodes (model sbt601 and sbw601) had been used with bodyguardian one and bodyguardian verite one devices.The initial reporter has attached for each patient an incident summary.In total, there were 10 reports covering a period from august 19 to september 28.All patients were prepared with different methods using water and soap, alcohol, alcohol wipes or saline wipes before applying the ecg electrodes.The patients were experiencing the "redness, itching, rash, blisters, welt, burning sensation on skin, peeling skin, burnt skin, bleeding, open sores." no medical intervention was necessary for any of these 10 incidents.No further information was provided.
|
|
Event Description
|
On (b)(6) 2020, we have been informed about 10 incidents with ecg electrodes.Monitoring ecg electrodes (model sbt601 and sbw601) had been used with bodyguardian one and bodyguardian verite one devices.The initial reporter has attached for each patient an incident summary.In total, there were 10 reports covering a period from (b)(6).All patients were prepared with different methods using water and soap, alcohol, alcohol wipes or saline wipes before applying the ecg electrodes.The patients were experiencing the "redness, itching, rash, blisters, welt, burning sensation on skin, peeling skin, burnt skin, bleeding, open sores." no medical intervention was necessary for any of these 10 incidents.No further information was provided.
|
|
Manufacturer Narrative
|
Retained samples of the concerened lots (191106-0225, 191015-0222, 191029-0155, 200121-0155, 200124-0157 and 200131-0251) of model sbt601 and model sbw601 have been inspected visually.Mechanical tests were performed on 3 retained samples.All tested samples were found to perform within limits.No faults could be detected.No information on the patient, skin type, state of skin, whether any medication was being taken, which might have a skin weakening effect, duration of use and details of the use was provided.We have requested further information several times but none was made available.The initial reporter stated that no further information has been made available despite repeated requests.We therefore consider the investigation closed.
|
|