The laboratory recently started reporting body fluids, and there was confusion in the laboratory regarding the xn units versus the lis units for the red blood cell body fluid (rbc-bf) parameter.The xn units were displayed as 10^3/ul, while the lis units were displayed as k/ul.Due to the unit difference, the laboratory adopted a manual process for reporting body fluid results for both the wbc-bf and rbc-bf parameters.During this manual entry process, the transcription error for the wbc-bf parameter occurred.The rbc-bf parameter was reported correctly.To resolve the issue, the laboratory changed the rbc-bf units on the xn analyzer.No analyzer deficiency was identified.
|
A patient underwent an unnecessary exploratory knee arthroscopy procedure due to an erroneously high white blood cell body fluid (wbc-bf) result.The patient's synovial fluid was analyzed on a hematology analyzer, and the analyzer generated an accurate result.The user manually entered the wbc-bf result in the laboratory information system (lis).When the user manually entered the result, the user unintentionally reported a value that was approximately 10 times greater than the actual result.The falsely increased wbc-bf result led to the unnecessary procedure.No patient harm was incurred due to the procedure.
|