The remaining shaft of the instrument has not been provided for evaluation.
Additionally, no radiographs have been provided for analysis.
While the reported complaint cannot be confirmed or properly evaluated at this time, given the age of the device (approximately 10 years), the failure reported is consistent with normal age and wear.
Review of labeling: warnings and precautions: seaspine instruments should only be used with seaspine implants.
Do not attempt to use with other competitive devices.
Avoid application of excessive stress on surgical instrumentation.
Carefully read and follow any package insert which accompanies the implants to be used with this instrumentation.
Instruments must be cleaned and decontaminated before they are returned to the manufacturer for any reason.
Carefully inspect all instruments prior to use.
Do not use an instrument that is severely marred or worn, or a cutting instrument with dull edges.
Note that at some time, instruments may wear out and should be replaced.
|
On (b)(6) 2020, the patient underwent spinal surgery consisting of seaspine's malibu pedicle screw system.
Seaspine was informed on 30 oct 2020 that the implant screwdriver broke while implanting a screw.
The driver tip was left in the screw in-situ.
No injury was reported, nor any additional medical or surgical treatment.
|