Model Number NCEUP4015X
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/29/2020
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
During a procedure, an attempt was made to use one nc euphora rx ptca balloon catheter to treat a very tortuous and calcified lesion located in the right coronary artery. It was reported that while attempting to place a stent in the rca, the nc euphora's balloon shaft broke off after multiple inflations. It was stated that all parts of the balloon were removed from the rca and the patient using the guidewire. No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameNC EUPHORA RX
Type of DeviceCATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX 22570
Manufacturer (Section G)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX 22570
Manufacturer Contact
toni o'doherty
parkmore business park west
galway 
EI  
091708734
MDR Report Key10913233
MDR Text Key218993181
Report Number9612164-2020-04649
Device Sequence Number1
Product Code LOX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141090
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation
Type of Report Initial,Followup,Followup
Report Date 02/10/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received11/27/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberNCEUP4015X
Device Catalogue NumberNCEUP4015X
Device Lot Number220243977
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received02/05/2021
Was Device Evaluated by Manufacturer? No Answer Provided
Date Device Manufactured06/19/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-