Model Number 4FC12
Device Problem Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/21/2020
Event Type  malfunction  
Manufacturer Narrative
Product event summary: the 4fc12 sheath with lot number 0009990642 was returned and analyzed. Visual inspection of the sheath showed a kink 2. 625 inches from the tip. The deflection worked as per specification. It was noted there could be steering difficulty due to a kink on the shaft. In conclusion, the sheath failed the returned product inspection due to a shaft kink. If information is provided in the future, a supplemental report will be issued.
 
Event Description
After a completed case, the sheath tested out of specification per the manufacturer's investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameFLEXCATH ADVANCE STEERABLE SHEATH
Type of DeviceCATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA H9R 5Z8
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key10920733
MDR Text Key218723998
Report Number3002648230-2020-00621
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 11/30/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received11/30/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date11/07/2021
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot Number0009990642
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/19/2020
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received11/09/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/08/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-