MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC DEPTH GAUGE FOR LOCKING SCREWS TO 100MM FOR IM NAILS; GAUGE, DEPTH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC DEPTH GAUGE FOR LOCKING SCREWS TO 100MM FOR IM NAILS; GAUGE, DEPTH Back to Search Results
Model Number 03.010.072
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Initial reporter is a synthes sales representative.Part 03.010.072, lot 3788247/1597033: manufacturing site: (b)(4).Release to warehouse date: july 22, 2011/november 29, 2006.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformance's were identified.A product investigation was completed: upon visual inspection, it was observed that the distal tip of the hook was bent and the slider assembly was missing the compression sleeve and the ball components.There were scratches and nicks on the device but have no impact on the device functionality.No other issues were identified with the returned device.A functional test was not performed as there was conclusive evidence that the returned device was missing components.No dimensional inspection can be performed due to post-manufacturing damage and missing components.Based on the date of manufacture, the current and manufactured revision of drawings were reviewed.The complaint condition was confirmed.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.The potential cause could be due to unintended forces applied to the device.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that on an unknown date, the depth gauge for locking screws to 100mm for nails would not slide.It is unknown if there was a surgical delay.It is unknown if procedure was successfully completed.There is no patient consequence.During the investigation by the manufacturer, it was noted the distal tip of the hook was bent and the slider assembly was missing the compression sleeve and the ball components.This report is for a depth gauge.This is report 1 of 1 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEPTH GAUGE FOR LOCKING SCREWS TO 100MM FOR IM NAILS
Type of Device
GAUGE, DEPTH
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
WERK BIO OBERDORF (CH)
eimattstrasse 3
oberdorf 4436
SZ   4436
Manufacturer Contact
kara ditty-bovard
1302 wright lane east
west chester, PA 19380
6107195000
MDR Report Key10921036
MDR Text Key218981302
Report Number2939274-2020-05360
Device Sequence Number1
Product Code HTJ
UDI-Device Identifier10886982067449
UDI-Public(01)10886982067449
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/30/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number03.010.072
Device Catalogue Number03.010.072
Device Lot Number3788247
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/11/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/17/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/22/2011
Is the Device Single Use? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-