MAUDE Adverse Event Report: WILLIAM COOK EUROPE UNKNOWN; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILLIAM COOK EUROPE UNKNOWN; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Catalog Number UNKNOWN
Device Problems Structural Problem (2506); Device Tipped Over (2589); Insufficient Information (3190)
Patient Problems Pain (1994); Swelling (2091); Perforation of Vessels (2135); No Information (3190)
Event Type  Injury  
Manufacturer Narrative
(b)(4).Catalog# is unknown but referred to as cook celect filter.Initial reporter occupation: non-healthcare professional.It has not been possible to investigate or evaluate this alleged event based on the limited information provided to date.Cook will reopen its investigation if further information is receiving warranting supplementation in accordance with 21 c.F.R.803.56.
 
Event Description
Description of event according to short form complaint filed: it is alleged that "[pt] received a cook celect filter on (b)(6) 2012.It is alleged that the [pt] was injured without further explanation.Hospital and medical records have been requested but not yet provided.".
 
Manufacturer Narrative
The following fields were updated per additional information received: a2, a4, b1, b5, b6, b7, h6.Additional information: investigation the following allegations have been investigated: vena cava perforation, tilt, swelling, pain, heaviness, physical limitations, requires compression stockings.Investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.Filter tilt has been reported.Potential causes may include filter placement in ivcs with diameters larger than those specified in these instructions for use; improper deployment; manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter); and (or) a failed retrieval attempt.Excessive filter tilt may contribute to difficult or failed retrieval; vena cava wall penetration/perforation; and (or) result in loss of filter efficiency.Potential adverse events that may occur include, but are not limited to, the following: unacceptable filter tilt.Unknown if the reported swelling, pain, heaviness, physical limitations, requires compression stockings are directly related to the filter and unable to identify a corresponding failure mode at this point in time.The catalog and lot numbers are unknown.The alleged celect is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56 this report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
Patient allegedly received an implant on (b)(6) 2012 via the right common femoral vein due to deep vein thrombosis (dvt).Patient is alleging tilt and vena cava perforation.Patient further alleges swelling in both legs, pain, heaviness, difficulty walking, required use of compression stockings report from computerized tomography (ct): "there is an ivc filter in place.The upper tip of the filter protrudes through the wall of the ivc laterally on the side consistent with a grade 2 perforation.All of the struts of the ivc protrude through the wall of the ivc consistent with grade 2 or 2-3 perforation.There is also angulation of the filter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
DA  4632
MDR Report Key10926813
MDR Text Key218914054
Report Number3002808486-2020-01114
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 12/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/01/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Date Manufacturer Received12/04/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
Patient Weight79
-
-